How the mounting of an accelerometer determines its usable frequency range.
Requirements
- Objective
- The mounting must not distort the frequency response and the measurement
- Mounting (from stiff to soft)
- Stud on a flat machined surface with coupling fluid, adhesive, wax or adhesive pad, magnet, hand-held probe
- Effect
- The softer the mounting, the lower the mounted resonance and thus the upper frequency limit
- Mass
- Keep the transducer mass small compared with the measuring point (mass loading)
- Cable
- Fix the cable close to the transducer to avoid cable whip and spurious signals